Type K
Type K Thermocouple
Type K Thermocouple
Data and Polynomial
- Description
Ni-Cr / Ni-Al
- Data source
National Institute of Science and Technology (NIST)
- Raw Data
- Octave format
- Curve fit
- mV -> temperature (C)
e is maximum error (in degrees C)
v1 andf v2 partition the domain into three segments, each with a polynomial fit
c is vector of coefficients: a10, a9, ... a0
Note: Each column of c corresponds to one range (<v1,v1 to v2, >v2)
e = 0.34668
v1 = 11.176
v2 = 22.691
c =
-2.4608e-07 -2.1453e-09
1.2296e-13
6.8153e-06 3.3858e-07
-4.9588e-11
-4.2163e-05 -2.3642e-05
8.7399e-09
-3.2201e-04 9.5925e-04
-8.9155e-07
3.5591e-03 -2.4958e-02
5.8569e-05
2.8323e-03 4.3295e-01
-2.5980e-03
-9.0658e-02 -5.0349e+00
7.8935e-02
1.2429e-01 3.8316e+01
-1.6195e+00
1.9257e-01 -1.7907e+02
2.1387e+01
2.4940e+01 4.7066e+02
-1.3999e+02
-7.2259e-01 -4.0375e+02
5.6264e+02
- temperature -> mV (range -127C-127C)
e is maximum error (in mV)
c is vector of coefficients: a10, a9, ... a0
e = 0.00067432
c =
-6.7683e-23
4.5979e-21
3.0278e-18
-1.8030e-16
-5.0020e-14
2.5828e-12
3.6705e-10
-1.3166e-07
2.6120e-05
3.9443e-02
-1.6551e-04